Importadora Biodinámica SpA · Scientific equipmentSantiago, Chile · biodynex.clEspañol
BiodynexRequest a quote
BIOBASE pulsed-vacuum vertical autoclave — 50-100 L, 105-134 °C
Autoclaves · Pulsed vacuum

BIOBASE pulsed-vacuum vertical autoclave

Family with vacuum extraction and steam fill cycles to improve penetration of the sterilizing medium into compatible loads.

  • Ejector vacuum system
  • Selectable programs
  • Movable probe for liquid control
  • Built-in printer
Configurations, certifications, power supply, accessories and availability are confirmed in the quote.
Specifications

Summary datasheet

Capacities50, 75 and 100 L
Working temperature105-134 °C
Design pressure-0.1/0.28 MPa
Working pressure0.23 MPa
Chamber materialS30408 (AISI 304)
Dimensions700 × 610 × 1070 mm (BKQ-B50V and BKQ-B75V); 700 × 610 × 1300 mm (BKQ-B100V)
AvailabilityImport to order / to be confirmed

What pulsed vacuum adds in this family

The complete cycle runs on its own and chains preheating, air replacement, pulsed exhaust, heating, sterilization and drying. The drying phase is the one a gravity autoclave does not have, and it is what allows the load to be taken out without residual moisture.

The catalog places it on textiles, rubber, instruments, laboratory utensils, culture media and unsealed liquids.

The 0.22 µm filter exists for a specific reason

Every time the cycle breaks the vacuum, outside air enters a chamber that has just been left sterile. The 0.22 µm filter sterilizes that air, so the load is not recontaminated by the very process that sterilized it.

It comes with the validation programs the gravity units do not have

On top of the usual cycles, this family includes a Bowie-Dick test and a vacuum test. Those are the two checks asked of a vacuum unit to show that it removes the air and that the chamber holds the vacuum.

Its altitude limit is the lowest in the range

As standard it takes up to 2000 m, against the 2700 m of the gravity families, and the manufacturer offers to customize it for the 2000 to 3000 m band. If the unit is going to a city above that level, it has to be stated in the order.